@inproceedings{3e811bcaad914f5da73bd3e878ec2226,
title = "Weight reduction of delta external fixation frames",
abstract = "The existing tibial delta external fixation frames consist of side bars and cross bars. In this work it is shown that the cross bars do not add appreciable strength to the frames. Therefore the removal of the cross bars of delta frame results in a new frame (V-frame) which is practically as strong as the conventional delta frame, but it is up to 35% lighter in weight.",
author = "J. Vossoughi and A. Poka and Burgess, {A. R.} and M. Bosse",
year = "1991",
month = dec,
day = "1",
language = "English (US)",
isbn = "0791808890",
series = "American Society of Mechanical Engineers, Bioengineering Division (Publication) BED",
publisher = "Publ by ASME",
pages = "151--154",
booktitle = "1991 Advances in Bioengineering",
note = "Winter Annual Meeting of the American Society of Mechanical Engineers ; Conference date: 01-12-1991 Through 06-12-1991",
}