TY - JOUR
T1 - Validation of a Wireless, Self-Application, Ambulatory Electroencephalographic Sleep Monitoring Device in Healthy Volunteers
AU - Finan, Patrick Hamilton
AU - Richards, Jessica M.
AU - Gamaldo, Charlene E.
AU - Han, Dingfen
AU - Leoutsakos, Jeannie Marie
AU - Salas, Rachel
AU - Irwin, Michael R.
AU - Smith, Michael T.
PY - 2016
Y1 - 2016
N2 - Study Objectives: To evaluate the validity of an ambulatory electroencephalographic (EEG) monitor for the estimation of sleep continuity and architecture in healthy adults. Methods: Healthy, good sleeping participants (n = 14) were fit with both an ambulatory EEG monitor (Sleep Profiler) and a full polysomnography (PSG) montage. EEG recordings were gathered from both devices on the same night, during which sleep was permitted uninterrupted for eight hours. The study was set in an inpatient clinical research suite. PSG and Sleep Profiler records were scored by a neurologist board certified in sleep medicine, blinded to record identification. Agreement between the scored PSG record, the physician-scored Sleep Profiler record, and the Sleep Profiler record scored by an automatic algorithm was evaluated for each sleep stage, with the PSG record serving as the reference. Results: Results indicated strong percent agreement across stages. Kappa was strongest for Stage N3 and REM. Specificity was high for all stages; sensitivity was low for Wake and Stage N1, and high for Stage N2, Stage N3, and REM. Agreement indices improved for the manually scored Sleep Profiler record relative to the autoscore record. Conclusions: Overall, the Sleep Profiler yields an EEG record with comparable sleep architecture estimates to PSG. Future studies should evaluate agreement between devices with a clinical sample that has greater periods of wake in order to better understand utility of this device for estimating sleep continuity indices, such as sleep onset latency and wake after sleep onset.
AB - Study Objectives: To evaluate the validity of an ambulatory electroencephalographic (EEG) monitor for the estimation of sleep continuity and architecture in healthy adults. Methods: Healthy, good sleeping participants (n = 14) were fit with both an ambulatory EEG monitor (Sleep Profiler) and a full polysomnography (PSG) montage. EEG recordings were gathered from both devices on the same night, during which sleep was permitted uninterrupted for eight hours. The study was set in an inpatient clinical research suite. PSG and Sleep Profiler records were scored by a neurologist board certified in sleep medicine, blinded to record identification. Agreement between the scored PSG record, the physician-scored Sleep Profiler record, and the Sleep Profiler record scored by an automatic algorithm was evaluated for each sleep stage, with the PSG record serving as the reference. Results: Results indicated strong percent agreement across stages. Kappa was strongest for Stage N3 and REM. Specificity was high for all stages; sensitivity was low for Wake and Stage N1, and high for Stage N2, Stage N3, and REM. Agreement indices improved for the manually scored Sleep Profiler record relative to the autoscore record. Conclusions: Overall, the Sleep Profiler yields an EEG record with comparable sleep architecture estimates to PSG. Future studies should evaluate agreement between devices with a clinical sample that has greater periods of wake in order to better understand utility of this device for estimating sleep continuity indices, such as sleep onset latency and wake after sleep onset.
KW - Ambulatory EEG
KW - Polysomnography
KW - Sleep
KW - Validation
UR - http://www.scopus.com/inward/record.url?scp=84998773691&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84998773691&partnerID=8YFLogxK
U2 - 10.5664/jcsm.6262
DO - 10.5664/jcsm.6262
M3 - Article
C2 - 27707438
AN - SCOPUS:84998773691
SN - 1550-9389
VL - 12
SP - 1443
EP - 1451
JO - Journal of Clinical Sleep Medicine
JF - Journal of Clinical Sleep Medicine
IS - 11
ER -