T1 - Subluxation of Suture-Fixated Posterior Chamber Intraocular Lenses. A Clinicopathologic Study
AU - Parekh, Parag
AU - Green, W. Richard
AU - Stark, Walter J.
AU - Akpek, Esen Karamursel
N1 - Funding Information:
Dr Green is supported in part by the Independent Order of Odd Fellows, Baltimore, Maryland. Dr Stark is supported in part by the Stavros Niarchos Research Fund, Baltimore, Maryland, and Helen and Raymond Kwok Research Fund, Baltimore, Maryland. Dr Akpek is supported in part by a William and Mary Greve Scholarship from Research to Prevent Blindness, New York, New York.
PY - 2007/2
Y1 - 2007/2
N2 - Purpose: To report the occurrence of subluxation of suture-fixated posterior chamber (PC) intraocular lenses (IOL) and elucidate the mechanisms involved. Design: Prospective clinicopathologic study. Participants: A single 10-0 Prolene suture explanted from a patient who experienced subluxation of his PC-IOL, 11.5 years after placement. Furthermore, multiple 10-0 Prolene sutures and PC-IOLs used for iris fixation were studied as controls. Methods: Scanning electron microscopy (SEM) was used to analyze the surface of the explanted suture. In addition, randomly selected 10-0 Prolene sutures cut with Vannas scissors and cut with the positioning holes of a randomly selected PC-IOL identical to that implanted in the patient's eye were examined as controls. Finally, the positioning holes of several randomly selected, iris-fixated PC-IOLs were studied using SEM with particular attention to surface quality and edge finish. Main Outcome Measures: Presence of any signs of suture degradation, the character of the cut edge of the suture, as well as the characteristics of the positioning holes of the PC-IOLs. Results: Scanning electron microscopy of the explanted suture revealed sharply cut edges, without significant degradation of the suture, and no intact loop. Scanning electron microscopy of the control suture cut with a PC-IOL demonstrated a similarly cut edge. The positioning holes of the examined PC-IOLs had a sharp edge, and some also had an imperfect finish. Conclusion: We conclude that the surface properties of the positioning holes lead to cutting of the suture, and subsequent subluxation of the PC-IOL.
AB - Purpose: To report the occurrence of subluxation of suture-fixated posterior chamber (PC) intraocular lenses (IOL) and elucidate the mechanisms involved. Design: Prospective clinicopathologic study. Participants: A single 10-0 Prolene suture explanted from a patient who experienced subluxation of his PC-IOL, 11.5 years after placement. Furthermore, multiple 10-0 Prolene sutures and PC-IOLs used for iris fixation were studied as controls. Methods: Scanning electron microscopy (SEM) was used to analyze the surface of the explanted suture. In addition, randomly selected 10-0 Prolene sutures cut with Vannas scissors and cut with the positioning holes of a randomly selected PC-IOL identical to that implanted in the patient's eye were examined as controls. Finally, the positioning holes of several randomly selected, iris-fixated PC-IOLs were studied using SEM with particular attention to surface quality and edge finish. Main Outcome Measures: Presence of any signs of suture degradation, the character of the cut edge of the suture, as well as the characteristics of the positioning holes of the PC-IOLs. Results: Scanning electron microscopy of the explanted suture revealed sharply cut edges, without significant degradation of the suture, and no intact loop. Scanning electron microscopy of the control suture cut with a PC-IOL demonstrated a similarly cut edge. The positioning holes of the examined PC-IOLs had a sharp edge, and some also had an imperfect finish. Conclusion: We conclude that the surface properties of the positioning holes lead to cutting of the suture, and subsequent subluxation of the PC-IOL.
UR - http://www.scopus.com/inward/record.url?scp=33846504373&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33846504373&partnerID=8YFLogxK
U2 - 10.1016/j.ophtha.2006.10.037
DO - 10.1016/j.ophtha.2006.10.037
M3 - Article
C2 - 17270672
AN - SCOPUS:33846504373
SN - 0161-6420
VL - 114
SP - 232
EP - 237
JO - Ophthalmology
JF - Ophthalmology
IS - 2
ER -