TY - JOUR
T1 - Sonolucent cranial implants
T2 - Cadaveric study and clinical findings supporting diagnostic and therapeutic transcranioplasty ultrasound
AU - Belzberg, Micah
AU - Shalom, Netanel Ben
AU - Yuhanna, Edward
AU - Manbachi, Amir
AU - Tekes, Aylin
AU - Huang, Judy
AU - Brem, Henry
AU - Gordon, Chad R.
N1 - Publisher Copyright:
© 2019 The Author(s).
PY - 2019
Y1 - 2019
N2 - Background: Previously, sonographic evaluation of the intracranial contents was limited to intraoperative use following bone flap removal, with placement of the probe directly on the cortical surface or through a transsulcal tubular retractor. Cranioplasty with sonolucent implants may represent a postoperative window into the brain by allowing ultrasound to serve as a novel bedside imaging modality. The potential sonolucency of various commonly used cranial implant types was examined in this study. Methods: A 3-phase study was comprised of cadaveric evaluation of transcranioplasty ultrasound (TCU) with cranioplasty implants of varying materials, intraoperative TCU during right-sided cranioplasty with clear implant made of poly-methyl-methacrylate (PMMA), and bedsideTCUon postoperative day 5 after cranioplasty. Results: The TCU through clear PMMA, polyether-ether-ketone, and opaque PMMA cranial implants revealed implant sonoluceny, in contrast to autologous bone and porous-polyethylene. Intraoperative ultrasound via the clear PMMA implant in a single patient revealed recognizable ventricular anatomy. Furthermore, postoperative bedside ultrasound in the same patient revealed comparable ventricular anatomy and a small epidural fluid collection corresponding to that visualized on an axial computed tomography scan. Conclusion: Sonolucent cranial implants, such as those made of clear PMMA, hold great promise for enhanced diagnostic and therapeutic applications previously limited by cranial bone. Furthermore, as functional cranial implants are manufactured with implantable devices housed within clear PMMA, the possibility of utilizing ultrasound for real-time surveillance of intracranial pathology becomes much more feasible.
AB - Background: Previously, sonographic evaluation of the intracranial contents was limited to intraoperative use following bone flap removal, with placement of the probe directly on the cortical surface or through a transsulcal tubular retractor. Cranioplasty with sonolucent implants may represent a postoperative window into the brain by allowing ultrasound to serve as a novel bedside imaging modality. The potential sonolucency of various commonly used cranial implant types was examined in this study. Methods: A 3-phase study was comprised of cadaveric evaluation of transcranioplasty ultrasound (TCU) with cranioplasty implants of varying materials, intraoperative TCU during right-sided cranioplasty with clear implant made of poly-methyl-methacrylate (PMMA), and bedsideTCUon postoperative day 5 after cranioplasty. Results: The TCU through clear PMMA, polyether-ether-ketone, and opaque PMMA cranial implants revealed implant sonoluceny, in contrast to autologous bone and porous-polyethylene. Intraoperative ultrasound via the clear PMMA implant in a single patient revealed recognizable ventricular anatomy. Furthermore, postoperative bedside ultrasound in the same patient revealed comparable ventricular anatomy and a small epidural fluid collection corresponding to that visualized on an axial computed tomography scan. Conclusion: Sonolucent cranial implants, such as those made of clear PMMA, hold great promise for enhanced diagnostic and therapeutic applications previously limited by cranial bone. Furthermore, as functional cranial implants are manufactured with implantable devices housed within clear PMMA, the possibility of utilizing ultrasound for real-time surveillance of intracranial pathology becomes much more feasible.
KW - Cranioplasty
KW - Implant
KW - Poly-methyl-methacrylate
KW - Sonolucent
KW - Ultrasound
UR - http://www.scopus.com/inward/record.url?scp=85068987649&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85068987649&partnerID=8YFLogxK
U2 - 10.1097/SCS.0000000000005454
DO - 10.1097/SCS.0000000000005454
M3 - Article
C2 - 31299743
AN - SCOPUS:85068987649
SN - 1049-2275
VL - 30
SP - 1456
EP - 1461
JO - Journal of Craniofacial Surgery
JF - Journal of Craniofacial Surgery
IS - 5
ER -