T1 - Sensing lead-related complications in patients with transvenous implantable cardioverter-defibrillators
AU - Lawton, Jennifer S.
AU - Ellenbogen, Kenneth A.
AU - Wood, Mark A.
AU - Stambler, Bruce S.
AU - Herre, John M.
AU - Nath, Sunil
AU - Bernstein, Robert C.
AU - DiMarco, John P.
AU - Haines, David E.
AU - Szentpetery, Szabolcs
AU - Baker, Lenox D.
AU - Damiano, Ralph J.
N1 - Copyright:
Copyright 2017 Elsevier B.V., All rights reserved.
PY - 1996
Y1 - 1996
N2 - The widespread use of the redesigned Endotak™ lead (CPI, St. Paul, Minnesota), which combines transvenous pacing, sensing, and defibrillation on a single transvenous lead in patients receiving transvenous implantable cardioverter-defibrillators (ICDs), has reduced morbidity and shortened length of hospital stay after ICD implantation. We describe the incidence and management of Endotak sensing lead-related failures in a series of 348 consecutive patients from 4 institutions who underwent implantation between 1990 and 1995. We retrospectively reviewed the databases for patients receiving an ICD with an Endotak lead for the incidence of lead-related sensing abnormalities. Ten patients (2.8%) with lead-related sensing abnormalities were detected at a mean of 15 ± 11 months after ICD implantation. Sensing abnormalities were detected in 6 patients after they received inappropriate shocks. Noise or oversensing was noted in 7 patients from interrogation of the devices' data logs. Eight patients had a new transvenous sensing lead placed, 1 patient had a new Endotak lead placed, and 1 had a chronic pacemaker sensing lead converted to function as a sensing lead. No further sensing problems were noted in 8 of 10 patients during a mean follow-up of 14 ± 8 months. The site of the sensing lead failure was localized to the subrectus pocket in 5 patients and to the clavicle-first rib area in 3 patients; it was undetermined and presumed to be in the clavicle-first rib area in the other 2 patients. One patient had late failure of the defibrillation lead. We conclude that Endotak sensing lead failure does not require insertion of a new Endotak lead, but can be managed with close follow-up and insertion of a new transvenous sensing lead. Endotak lead fractures are frequently localized to the ICD pocket.
AB - The widespread use of the redesigned Endotak™ lead (CPI, St. Paul, Minnesota), which combines transvenous pacing, sensing, and defibrillation on a single transvenous lead in patients receiving transvenous implantable cardioverter-defibrillators (ICDs), has reduced morbidity and shortened length of hospital stay after ICD implantation. We describe the incidence and management of Endotak sensing lead-related failures in a series of 348 consecutive patients from 4 institutions who underwent implantation between 1990 and 1995. We retrospectively reviewed the databases for patients receiving an ICD with an Endotak lead for the incidence of lead-related sensing abnormalities. Ten patients (2.8%) with lead-related sensing abnormalities were detected at a mean of 15 ± 11 months after ICD implantation. Sensing abnormalities were detected in 6 patients after they received inappropriate shocks. Noise or oversensing was noted in 7 patients from interrogation of the devices' data logs. Eight patients had a new transvenous sensing lead placed, 1 patient had a new Endotak lead placed, and 1 had a chronic pacemaker sensing lead converted to function as a sensing lead. No further sensing problems were noted in 8 of 10 patients during a mean follow-up of 14 ± 8 months. The site of the sensing lead failure was localized to the subrectus pocket in 5 patients and to the clavicle-first rib area in 3 patients; it was undetermined and presumed to be in the clavicle-first rib area in the other 2 patients. One patient had late failure of the defibrillation lead. We conclude that Endotak sensing lead failure does not require insertion of a new Endotak lead, but can be managed with close follow-up and insertion of a new transvenous sensing lead. Endotak lead fractures are frequently localized to the ICD pocket.
UR - http://www.scopus.com/inward/record.url?scp=10644244566&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=10644244566&partnerID=8YFLogxK
U2 - 10.1016/S0002-9149(96)00387-6
DO - 10.1016/S0002-9149(96)00387-6
M3 - Article
C2 - 8831398
AN - SCOPUS:10644244566
SN - 0002-9149
VL - 78
SP - 647
EP - 651
JO - American Journal of Cardiology
JF - American Journal of Cardiology
IS - 6
ER -