@inproceedings{59c622bb0f8348a69e245ff40d3512ed,
title = "Pulse pileup statistics for energy sensitive photon counting detectors with pulse height analysis",
abstract = "Photon counting statistics with pulse pileup effects (PPE) have been investigated for detection schemes with two assumptions: a fixed deadtime (i.e., non-paralyable and paralyzable detection schemes) and a delta pulse shape. Analytical expressions have been developed which shed light on interesting findings: (1) the variance becomes smaller than mean with PPE; and (2) the variance-to-mean ratio (VMR) of narrow energy windows remains very close to 1. In this study, we experimentally investigate VMR with PPE with a variable deadtime (i.e., pulse height analysis) and bipolar and unipolar pulse shapes with a long tail in addition to the above cases. We will use Monte Carlo simulation for a systematic study and a physical photon counting detector to confirm the results.",
keywords = "Computed tomography",
author = "Katsuyuki Taguchi and Somesh Srivastava and Qiulin Tang and Caffo, {Brian S.} and Iwanczyk, {Jan S.} and Hartsough, {Neal E.} and Barber, {William C.} and Jochen Cammin",
year = "2012",
doi = "10.1117/12.911365",
language = "English (US)",
isbn = "9780819489623",
series = "Progress in Biomedical Optics and Imaging - Proceedings of SPIE",
booktitle = "Medical Imaging 2012",
note = "Medical Imaging 2012: Physics of Medical Imaging ; Conference date: 05-02-2012 Through 08-02-2012",
}