TY - JOUR
T1 - PREVENtion of HeartMate II Pump Thrombosis Through Clinical Management
T2 - The PREVENT multi-center study
AU - PREVENT Study Investigators
AU - Maltais, Simon
AU - Kilic, Ahmet
AU - Nathan, Sriram
AU - Keebler, Mary
AU - Emani, Sitaramesh
AU - Ransom, John
AU - Katz, Jason N.
AU - Sheridan, Brett
AU - Brieke, Andreas
AU - Egnaczyk, Gregory
AU - Entwistle, John W.
AU - Adamson, Robert
AU - Stulak, John
AU - Uriel, Nir
AU - O'Connell, John B.
AU - Farrar, David J.
AU - Sundareswaran, Kartik S.
AU - Gregoric, Igor
N1 - Publisher Copyright:
© 2016 The Authors
PY - 2017/1/1
Y1 - 2017/1/1
N2 - Background Recommended structured clinical practices including implant technique, anti-coagulation strategy, and pump speed management (PREVENT [PREVENtion of HeartMate II Pump Thrombosis Through Clinical Management] recommendations) were developed to address risk of early (<3 months) pump thrombosis (PT) risk with HeartMate II (HMII; St. Jude Medical, Inc. [Thoratec Corporation], Pleasanton, CA). We prospectively assessed the HMII PT rate in the current era when participating centers adhered to the PREVENT recommendations. Methods PREVENT was a prospective, multi-center, single-arm, non-randomized study of 300 patients implanted with HMII at 24 participating sites. Confirmed PT (any suspected PT confirmed visually and/or adjudicated by an independent assessor) was evaluated at 3 months (primary end-point) and at 6 months after implantation. Results The population included 83% men (age 57 years ± 13), 78% destination therapy, and 83% Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) Profile 1–3. Primary end-point analysis showed a confirmed PT of 2.9% at 3 months and 4.8% at 6 months. Adherence to key recommendations included 78% to surgical recommendations, 95% to heparin bridging, and 79% to pump speeds ≥9,000 RPMs (92% >8,600 RPMs). Full adherence to implant techniques, heparin bridging, and pump speeds ≥9,000 RPMs resulted in a significantly lower risk of PT (1.9% vs 8.9%; p < 0.01) and lower composite risk of suspected thrombosis, hemolysis, and ischemic stroke (5.7% vs 17.7%; p < 0.01) at 6 months. Conclusions Adoption of all components of a structured surgical implant technique and clinical management strategy (PREVENT recommendations) is associated with low rates of confirmed PT.
AB - Background Recommended structured clinical practices including implant technique, anti-coagulation strategy, and pump speed management (PREVENT [PREVENtion of HeartMate II Pump Thrombosis Through Clinical Management] recommendations) were developed to address risk of early (<3 months) pump thrombosis (PT) risk with HeartMate II (HMII; St. Jude Medical, Inc. [Thoratec Corporation], Pleasanton, CA). We prospectively assessed the HMII PT rate in the current era when participating centers adhered to the PREVENT recommendations. Methods PREVENT was a prospective, multi-center, single-arm, non-randomized study of 300 patients implanted with HMII at 24 participating sites. Confirmed PT (any suspected PT confirmed visually and/or adjudicated by an independent assessor) was evaluated at 3 months (primary end-point) and at 6 months after implantation. Results The population included 83% men (age 57 years ± 13), 78% destination therapy, and 83% Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) Profile 1–3. Primary end-point analysis showed a confirmed PT of 2.9% at 3 months and 4.8% at 6 months. Adherence to key recommendations included 78% to surgical recommendations, 95% to heparin bridging, and 79% to pump speeds ≥9,000 RPMs (92% >8,600 RPMs). Full adherence to implant techniques, heparin bridging, and pump speeds ≥9,000 RPMs resulted in a significantly lower risk of PT (1.9% vs 8.9%; p < 0.01) and lower composite risk of suspected thrombosis, hemolysis, and ischemic stroke (5.7% vs 17.7%; p < 0.01) at 6 months. Conclusions Adoption of all components of a structured surgical implant technique and clinical management strategy (PREVENT recommendations) is associated with low rates of confirmed PT.
KW - HeartMate II
KW - best practices
KW - continuous-flow left ventricular assist device (CF-LVAD)
KW - prevention
KW - pump thrombosis
UR - http://www.scopus.com/inward/record.url?scp=85007012266&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85007012266&partnerID=8YFLogxK
U2 - 10.1016/j.healun.2016.10.001
DO - 10.1016/j.healun.2016.10.001
M3 - Article
C2 - 27865732
AN - SCOPUS:85007012266
SN - 1053-2498
VL - 36
SP - 1
EP - 12
JO - Journal of Heart and Lung Transplantation
JF - Journal of Heart and Lung Transplantation
IS - 1
ER -