title = "Pixel-to-Pixel Variation Correction Using Cylindrical Phantoms in Photon-Counting CT: Total Count Results",
abstract = "Photon-counting detectors (PCDs) in computed tomography (CT) offer advantages over conventional detectors, such as improved spatial resolution and material discrimination. PCDs, however, exhibit greater pixel-to-pixel variability, leading to ring artifacts. This study proposes a calibration method using cylindrical water phantoms to correct these variations. The method involves two steps: acquiring calibration data by positioning phantoms at various offsets from the iso-center and estimating {\textquotedblleft}good pixel{\textquotedblright} responses to obtain pixel-to-pixel variation correction coefficients. Evaluations using a prototype PCD-based CT scanner showed a significant reduction in ring artifacts and improved image quality in test phantom data, validating the method's effectiveness. This approach enhances PCD-based CT systems and paves the way for future improvements in CT imaging.",
keywords = "Pixel-to-pixel variation, calibration, computed tomography, photon-counting detector, ring artifact",