TY - JOUR
T1 - MR-guided endovascular interventions
T2 - Device visualization, tracking, navigation, clinical applications, and safety aspects
AU - Wacker, Frank K.
AU - Hillenbrand, Claudia M.
AU - Duerk, Jeffrey L.
AU - Lewin, Jonathan S.
PY - 2005/8/1
Y1 - 2005/8/1
N2 - Reliable visualization and tracking are essential for guiding endovascular devices within blood vessels. The most commonly used methods are susceptibility artifact-based tracking that relies on the artifact created within the image by the device and microcoil- or antenna-based tracking that uses the high signal generated by small MR endovascular receive coils when the transmit coil emits a nonselective radiofrequency pulse. To date, the use of endovascular MR-guidance techniques has primarily been confined to animal experiments. There are only a few reports on MR-guided endovascular applications in patients. Therefore, access to the patient within the scanner, dedicated devices, and safety issues remain major challenges. To face these challenges, attention from all radiologists, especially interventional radiologists, is required to make MR-guided endovascular procedures a clinical reality.
AB - Reliable visualization and tracking are essential for guiding endovascular devices within blood vessels. The most commonly used methods are susceptibility artifact-based tracking that relies on the artifact created within the image by the device and microcoil- or antenna-based tracking that uses the high signal generated by small MR endovascular receive coils when the transmit coil emits a nonselective radiofrequency pulse. To date, the use of endovascular MR-guidance techniques has primarily been confined to animal experiments. There are only a few reports on MR-guided endovascular applications in patients. Therefore, access to the patient within the scanner, dedicated devices, and safety issues remain major challenges. To face these challenges, attention from all radiologists, especially interventional radiologists, is required to make MR-guided endovascular procedures a clinical reality.
UR - http://www.scopus.com/inward/record.url?scp=23444440180&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=23444440180&partnerID=8YFLogxK
U2 - 10.1016/j.mric.2005.04.004
DO - 10.1016/j.mric.2005.04.004
M3 - Review article
C2 - 16084411
AN - SCOPUS:23444440180
SN - 1064-9689
VL - 13
SP - 431
EP - 439
JO - Magnetic Resonance Imaging Clinics of North America
JF - Magnetic Resonance Imaging Clinics of North America
IS - 3
ER -