Abstract
This study quantified the torsional resistance provided by locking plates and angled blade plates used to stabilize proximal humeral fractures. Three-part proximal humeral fractures were created in 6 pairs of cadaveric humeri. One specimen of each pair was reconstructed with a proximal humeral locking plate, whereas the other specimen was reconstructed with an angled blade plate. An external rotation torque, varying from 0 to 5 N-m, was applied to the humeral head until the head rotated 30° or 10,000 loading cycles were applied. The mean initial torsional stiffness was significantly larger for the locking plates (0.99 N-m/degree) than for the blade plates (0.59 N-m/degree). For each pair, the maximum rotation was larger for the blade plate than for the locking plate. For this in vitro model of a reconstructed 3-part proximal humeral fracture, the locking plate provided better torsional fatigue resistance and stiffness than the blade plate.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 239-243 |
| Number of pages | 5 |
| Journal | Journal of Shoulder and Elbow Surgery |
| Volume | 15 |
| Issue number | 2 |
| DOIs | |
| State | Published - Mar 2006 |
| Externally published | Yes |
ASJC Scopus subject areas
- Surgery
- Orthopedics and Sports Medicine
Fingerprint
Dive into the research topics of 'Locking plates improve torsional resistance in the stabilization of three-part proximal humeral fractures'. Together they form a unique fingerprint.Cite this