TY - JOUR
T1 - Localization accuracy and immobilization effectiveness of a stereotactic body frame for a variety of treatment sites
AU - Foster, Ryan
AU - Meyer, Jeffrey
AU - Iyengar, Puneeth
AU - Pistenmaa, David
AU - Timmerman, Robert
AU - Choy, Hak
AU - Solberg, Timothy
PY - 2013/12/1
Y1 - 2013/12/1
N2 - Purpose The purpose of this study was to analyze the pretreatment setup errors and intrafraction motion using cone beam computed tomography (CBCT) for stereotactic body radiation therapy patients immobilized and localized with a stereotactic body frame for a variety of treatment sites. Methods and Materials Localization errors were recorded for patients receiving SBRT for 141 lung, 29 liver, 48 prostate, and 45 spine tumors representing 1005 total localization sessions. All patients were treated in a stereotactic body frame with a large custom-molded vacuum pillow. Patients were first localized to the frame using tattoos placed during simulation. Subsequently, the frame was aligned to the room lasers according to the stereotactic coordinates determined from the treatment plan. Every patient received a pretreatment and an intrafraction CBCT. Abdominal compression was used for all liver patients and for approximately 40% of the lung patients to reduce tumor motion due to respiration. Results The mean ± standard deviation pretreatment setup errors from all localizations were -2.44 ± 3.85, 1.31 ± 5.84, and 0.11 ± 3.76 mm in the anteroposterior, superoinferior, and lateral directions, respectively. The mean pretreatment localization results among all treatment sites were not significantly different (F test, P<.05). For all treatment sites, the mean ± standard deviation intrafraction shifts were 0.33 ± 1.34, 0.15 ± 1.45, and -0.02 ± 1.17 mm in the anteroposterior, superoinferior, and lateral directions, respectively. The mean unidimensional intrafraction shifts were statistically different for several of the comparisons (P<.05) as assessed by the Tukey-Kramer test. Conclusions Despite the varied tumor locations, the pretreatment mean localization errors for all sites were found to be consistent among the treatment sites and not significantly different, indicating that the body frame is a suitable immobilization and localization device for a variety of tumor sites. Our pretreatment localization errors and intrafraction shifts compare favorably with those reported in other studies using different types of immobilization devices.
AB - Purpose The purpose of this study was to analyze the pretreatment setup errors and intrafraction motion using cone beam computed tomography (CBCT) for stereotactic body radiation therapy patients immobilized and localized with a stereotactic body frame for a variety of treatment sites. Methods and Materials Localization errors were recorded for patients receiving SBRT for 141 lung, 29 liver, 48 prostate, and 45 spine tumors representing 1005 total localization sessions. All patients were treated in a stereotactic body frame with a large custom-molded vacuum pillow. Patients were first localized to the frame using tattoos placed during simulation. Subsequently, the frame was aligned to the room lasers according to the stereotactic coordinates determined from the treatment plan. Every patient received a pretreatment and an intrafraction CBCT. Abdominal compression was used for all liver patients and for approximately 40% of the lung patients to reduce tumor motion due to respiration. Results The mean ± standard deviation pretreatment setup errors from all localizations were -2.44 ± 3.85, 1.31 ± 5.84, and 0.11 ± 3.76 mm in the anteroposterior, superoinferior, and lateral directions, respectively. The mean pretreatment localization results among all treatment sites were not significantly different (F test, P<.05). For all treatment sites, the mean ± standard deviation intrafraction shifts were 0.33 ± 1.34, 0.15 ± 1.45, and -0.02 ± 1.17 mm in the anteroposterior, superoinferior, and lateral directions, respectively. The mean unidimensional intrafraction shifts were statistically different for several of the comparisons (P<.05) as assessed by the Tukey-Kramer test. Conclusions Despite the varied tumor locations, the pretreatment mean localization errors for all sites were found to be consistent among the treatment sites and not significantly different, indicating that the body frame is a suitable immobilization and localization device for a variety of tumor sites. Our pretreatment localization errors and intrafraction shifts compare favorably with those reported in other studies using different types of immobilization devices.
UR - http://www.scopus.com/inward/record.url?scp=84888300018&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84888300018&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2013.09.020
DO - 10.1016/j.ijrobp.2013.09.020
M3 - Article
C2 - 24351410
AN - SCOPUS:84888300018
SN - 0360-3016
VL - 87
SP - 911
EP - 916
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 5
ER -