Abstract
A reversed-phase ion-pair liquid chromatographic method with UV detection at 210 nm is described for the separation of lincomycin from 7-epilincomycin and lincomycin B. The method utilizes a base-deactivated Supelcosil LC-ABZ, C12-18, 5 μm, 250x4.6 mm I.D. column maintained at 45°C. The mobile phase consists of 2.25% (v/v) acetonitrile, 5% (v/v) phosphate buffer (2.72%, m/v KH2PO4 adjusted to pH 5.0 with 3.48%, m/v K2HPO4), 0.067% (v/v) methanesulfonic acid and water to 100%. The flow-rate is 1.0 ml/min. The separation of lincomycin from 7-epilincomycin is only possible on base-deactivated (BDS) columns. Other octadecylsilyl columns examined showed insufficient selectivity. The method was also tested on other BDS columns (Spherisorb S5-ODS-B, 5 μm, Hypersil BDS, C18, 5 μm and Supelcosil LC-ABZ, C18, 5 μm, all 250x4.6 mm I.D.) and showed good robustness. Robustness was further evaluated by performing a full-factorial design experiment. The method showed good selectivity, repeatability, linearity and sensitivity. It is also suitable for analysis of lincomycin formulations. Copyright (C) 1998 Elsevier Science B.V.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 161-166 |
| Number of pages | 6 |
| Journal | Journal of Chromatography A |
| Volume | 829 |
| Issue number | 1-2 |
| DOIs | |
| State | Published - 1998 |
Keywords
- Antibiotics
- Lincomycins
- Pharmaceutical analysis
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Organic Chemistry
Fingerprint
Dive into the research topics of 'Liquid chromatographic method for separation of lincomycin from its related substances'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS