Abstract
Two procedures for recording spectroscopic images with pure absorption mode lineshapes using phase-encoding pulsed field gradients are discussed and compared experimentally. Using a small phantom sample, both methods are shown to give significant improvements compared to the conventional absolute value display. It is found that the method based on the use of a symmetric data set is preferable to that based on a reversed precession experiment. To demonstrate its applicability to in vivo31P NMR, a surface coil placed over the scalp of a rabbit was used to record a one-dimensional phase-encoded spectrum. While the absolute value display of this data set contains overlap between the low-field ATP and the phosphocreatine resonances, these signals are completely resolved in the phase-sensitive spectrum.
Original language | English (US) |
---|---|
Pages (from-to) | 467-473 |
Number of pages | 7 |
Journal | Journal of Magnetic Resonance (1969) |
Volume | 75 |
Issue number | 3 |
DOIs | |
State | Published - Dec 1987 |
Externally published | Yes |