Skip to main navigation Skip to search Skip to main content

Fatigue properties of the Hoffmann-Vidal external fixation apparatus

Research output: Contribution to journalArticlepeer-review

Abstract

Fatigue tests of the standard Hoffmann-Vidal (H-V) external fixator were performed. After 2,500,000 cycles, fractures occurred in the post attached to the pin bracket. The screws on the pin fixation bracket tend to become loose after cyclic loading test, and periodic lightening of all the fixation screws on a H-V apparatus is recommended under in vivo application. In repeated lightening of individual joints, screws on the pin brackets and adjustable connecting rods fractured. Deformation of the retaining set screws in the universal ball joints caused joint stiffness reduction. Severe abrasive wear of the joint surfaces was observed after repeated lightening. It is concluded that this apparatus may be safely used for four to five six-month clinical applications, provided that the critical parts are replaced between sequential applications. Suggestions for clinical management and recommendations for component modifications are made.

Original languageEnglish (US)
Pages (from-to)443-456
Number of pages14
JournalOrthopedics
Volume7
Issue number3
DOIs
StatePublished - 1984
Externally publishedYes

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'Fatigue properties of the Hoffmann-Vidal external fixation apparatus'. Together they form a unique fingerprint.

Cite this