Evaluation of parallel and fan-beam data acquisition geometries and strategies for myocardial SPECT imaging

Yujin Qi, Benjamin M.W. Tsui, Karen L. Gilland, Eric C. Frey, Grant T. Gullberg

Research output: Contribution to journalArticlepeer-review

5 Scopus citations

Abstract

This study evaluates myocardial SPECT images obtained from parallel-hole (PH) and fan-beam (FB) collimator geometries using both circular-orbit (CO) and noncircular-orbit (NCO) acquisitions. A newly developed 4-D NURBS-based cardiac-torso (NCAT) phantom was used to simulate the 99mTc-sestamibi uptakes in human torso with myocardial defects in the left ventricular (LV) wall. Two phantoms were generated to simulate patients with thick and thin body builds. Projection data including the effects of attenuation, collimator-detector response and scatter were generated using SIMSET Monte Carlo simulations. A large number of photon histories were generated such that the projection data were close to noise free. Poisson noise fluctuations were then added to simulate the count densities found in clinical data. Noise-free and noisy projection data were reconstructed using the iterative OS-EM reconstruction algorithm with attenuation compensation. The reconstructed images from noisy projection data show that the noise levels are lower for the FB as compared to the PH collimator due to increase in detected counts. The NCO acquisition method provides slightly better resolution and small improvement in defect contrast as compared to the CO acquisition method in noise-free reconstructed images. Despite lower projection counts the NCO shows the same noise level as the CO in the attenuation corrected reconstruction images. The results from the channelized Hotelling observer (CHO) study show that FB collimator is superior to PH collimator in myocardial defect detection, but the NCO shows no statistical significant difference from the CO for either PH or FB collimator. In conclusion, our results indicate that data acquisition using NCO makes a very small improvement in the resolution over CO for myocardial SPECT imaging. This small improvement does not make a significant difference on myocardial defect detection. However, an FB collimator provides better defect detection than a PH collimator with similar spatial resolution for myocardial SPECT imaging.

Original languageEnglish (US)
Pages (from-to)667-672
Number of pages6
JournalIEEE Transactions on Nuclear Science
Volume51
Issue number3 II
DOIs
StatePublished - Jun 2004

Keywords

  • Channelized Hotelling observer (CHO)
  • Fan beam
  • ROC analysis
  • SPECT

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Nuclear Energy and Engineering
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Evaluation of parallel and fan-beam data acquisition geometries and strategies for myocardial SPECT imaging'. Together they form a unique fingerprint.

Cite this