Dosimetric Validation for Prospective Clinical Trial of GRID Collimator-Based Spatially Fractionated Radiation Therapy: Dose Metrics Consistency and Heterogeneous Pattern Reproducibility

Hualin Zhang, Lijun Ma, Andrew Lim, Jason Ye, Lauren Lukas, Heng Li, Nina A. Mayr, Eric Lin Chang

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: Dose heterogeneity within a tumor target is likely responsible for the biologic effects and local tumor control from spatially fractionated radiation therapy (SFRT). This study used a commercially available GRID-pattern dose mudulated nonuniform radiation therapy (GRID) collimator to assess the interplan variability of heterogeneity dose metrics in patients with various bulky tumor sizes and depths. Methods and Materials: The 3-dimensional heterogeneity metrics of 14 bulky tumors, ranging from 155 to 2161 cm3 in volume, 6 to 23 cm in equivalent diameter, and 3 to 13 cm in depth, and treated with GRID collimator-based SFRT were studied. A prescription dose of 15 Gy was given at the tumor center with 6 MV photons. The dose-volume histogram indices, dose heterogeneity parameters, and peak/valley dose ratios were derived; the equivalent uniform doses of cancer cells with various radiosensitivities in each plan were estimated. To account for the spatial fractionation, high dose core number density of the tumor target was defined and calculated. Results: Among 14 plans, the dose-volume histogram indices D5, D10, D50, D90, and D95 (doses covering 5%, 10%, 50%, 90%, and 95% of the target volume) were found within 10% variation. The dose ratio of D10/D90 also showed a moderate consistency (range, 3.9-5.0; mean, 4.4). The equivalent uniform doses were consistent, ranging from 4.3 to 5.5 Gy, mean 4.6 Gy, for radiosensitive cancer cells and from 5.8 to 6.9 Gy, mean 6.2 Gy, for radioresistant cancer cells. The high dose core number density was within 20% among all plans. Conclusions: GRID collimator-based SFRT delivers a consistent heterogeneity dose distribution and high dose core density across bulky tumor plans. The interplan reproducibility and simplicity of GRID therapy may be useful for certain clinical indications and interinstitutional clinical trial design, and its heterogeneity metrics may help guide multileaf-collimator-based SFRT planning to achieve similar or further optimized dose distributions.

Original languageEnglish (US)
Pages (from-to)565-573
Number of pages9
JournalInternational Journal of Radiation Oncology Biology Physics
Volume118
Issue number2
DOIs
StatePublished - Feb 1 2024

ASJC Scopus subject areas

  • Radiation
  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Cancer Research

Fingerprint

Dive into the research topics of 'Dosimetric Validation for Prospective Clinical Trial of GRID Collimator-Based Spatially Fractionated Radiation Therapy: Dose Metrics Consistency and Heterogeneous Pattern Reproducibility'. Together they form a unique fingerprint.