TY - JOUR
T1 - Development of a carbon-coated, central-hinging, bileaflet valve
AU - Gott, Vincent L.
AU - Daggett, Ronald L.
AU - Young, William P.
PY - 1989/9
Y1 - 1989/9
N2 - Between April 1963 and January 1966, 86 patients at the University of Wisconsin Hospital underwent aortic or mitral valve replacement, or both, with a carbon-coated, central-hinging, bileaflet valve. A 25-year follow-up has been obtained on 43 of these patients discharged with an aortic prothesis, 13 patients discharged with a mitral prostheses, and 2 patients discharged with double prostheses. The mean implantation time was 7.2 years, 9.0 years, and 9.5 years, respectively. The longest aortic valve implantation time was 24.2 years in a patient who had her bileaflet valve prophylactically replaced, and the longest mitral implantation time is 24 years in a patient who is doing well with her original prosthesis. We are not aware of any patient among the approximately 700 receiving this valve around the world who has developed fatigue-failure of the silicone-impregnated Teflon fabric leaflet. This valve has demonstrated unexpected durability and has provided some design and biomaterial concepts that are used in a number of current prosthetic valves.
AB - Between April 1963 and January 1966, 86 patients at the University of Wisconsin Hospital underwent aortic or mitral valve replacement, or both, with a carbon-coated, central-hinging, bileaflet valve. A 25-year follow-up has been obtained on 43 of these patients discharged with an aortic prothesis, 13 patients discharged with a mitral prostheses, and 2 patients discharged with double prostheses. The mean implantation time was 7.2 years, 9.0 years, and 9.5 years, respectively. The longest aortic valve implantation time was 24.2 years in a patient who had her bileaflet valve prophylactically replaced, and the longest mitral implantation time is 24 years in a patient who is doing well with her original prosthesis. We are not aware of any patient among the approximately 700 receiving this valve around the world who has developed fatigue-failure of the silicone-impregnated Teflon fabric leaflet. This valve has demonstrated unexpected durability and has provided some design and biomaterial concepts that are used in a number of current prosthetic valves.
UR - http://www.scopus.com/inward/record.url?scp=0024341912&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0024341912&partnerID=8YFLogxK
U2 - 10.1016/0003-4975(89)90627-9
DO - 10.1016/0003-4975(89)90627-9
M3 - Article
C2 - 2673095
AN - SCOPUS:0024341912
SN - 0003-4975
VL - 48
SP - S28-S30
JO - The Annals of thoracic surgery
JF - The Annals of thoracic surgery
IS - 3 SUPPL.
ER -