Current pediatric administered activity guidelines for 99mTc-DMSA SPECT based on patient weight do not provide the same task-based image quality

Ye Li, Shannon O'Reilly, Donika Plyku, S. Ted Treves, Frederic Fahey, Yong Du, Xinhua Cao, Briana Sexton-Stallone, Justin Brown, George Sgouros, Wesley E. Bolch, Eric C. Frey

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Purpose: In the current clinical practice, administered activity (AA) for pediatric molecular imaging is often based on the North American expert consensus guidelines or the European Association of Nuclear Medicine dosage card, both of which were developed based on the best clinical practice. These guidelines were not formulated using a rigorous evaluation of diagnostic image quality (IQ) relative to AA. In the guidelines, AA is determined by a weight-based scaling of the adult AA, along with minimum and maximum AA constraints. In this study, we use task-based IQ assessment methods to rigorously evaluate the efficacy of weight-based scaling in equalizing IQ using a population of pediatric patients of different ages and body weights. Methods: A previously developed projection image database was used. We measured task-based IQ, with respect to the detection of a renal functional defect at six different AA levels (AA relative to the AA obtained from the guidelines). IQ was assessed using an anthropomorphic model observer. Receiver-operating characteristics (ROC) analysis was applied; the area under the ROC curve (AUC) served as a figure-of-merit for task performance. In addition, we investigated patient girth (circumference) as a potential improved predictor of the IQ. Results: The data demonstrate a monotonic and modestly saturating increase in AUC with increasing AA, indicating that defect detectability was limited by quantum noise and the effects of object variability were modest over the range of AA levels studied. The AA for a given value of the AUC increased with increasing age. The AUC vs AA plots for all the patient ages indicate that, for the current guidelines, the newborn and 10- and 15-yr phantoms had similar IQ for the same AA suggested by the North American expert consensus guidelines, but the 5- and 1-yr phantoms had lower IQ. The results also showed that girth has a stronger correlation with the needed AA to provide a constant AUC for 99mTc-DMSA renal SPECT. Conclusions: The results suggest that (a) weight-based scaling is not sufficient to equalize task-based IQ for patients of different weights in pediatric 99mTc-DMSA renal SPECT; and (b) patient girth should be considered instead of weight in developing new administration guidelines for pediatric patients.

Original languageEnglish (US)
Pages (from-to)4847-4856
Number of pages10
JournalMedical physics
Volume46
Issue number11
DOIs
StatePublished - Nov 1 2019

Keywords

  • DMSA
  • SPECT
  • administered activity guidelines
  • dose reduction/optimization
  • pediatric imaging
  • task-based image quality

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Current pediatric administered activity guidelines for 99mTc-DMSA SPECT based on patient weight do not provide the same task-based image quality'. Together they form a unique fingerprint.

Cite this