TY - JOUR
T1 - Complex retrieval of fractured, embedded, and penetrating inferior vena cava filters
T2 - A prospective study with histologic and electron microscopic analysis
AU - Kuo, William T.
AU - Robertson, Scott W.
AU - Odegaard, Justin I.
AU - Hofmann, Lawrence V.
PY - 2013/5
Y1 - 2013/5
N2 - Purpose: To evaluate clinical outcomes, characterize adherent tissue, and analyze inferior vena cava (IVC) filter fractures in patients undergoing complex retrieval for management of filter-related complications. To elucidate mechanisms of filter fracture by radiographic and electron microscopic (EM) evaluation. Materials and Methods: Over 2.5 years, 50 consecutive patients with fractured and/or penetrating filter components were prospectively enrolled into a single-center study. There were 19 men and 31 women (mean age, 42 y; range, 15-73 y). All patients underwent complex filter retrieval after failure of standard methods, and retrieval indications along with resultant clinical outcomes were evaluated. Specimens with adherent tissue underwent histologic analysis, and all fractured components were studied with EM. Results: Retrieval was successful in all 50 cases (mean implantation, 815 d; range, 20-2,599 d) among the following filters: G2X (n = 23),G2 (n = 9), Eclipse (n = 3), Recovery (n = 4), ALN (n = 1), Celect (n = 7), OptEase (n = 2), and Simon Nitinol (n = 1). Mean indwell time in fractured filters (n = 31) was 1,082 days, versus 408 days in nonfractured filters (n = 19; P =.00169). Neointimal hyperplasia/fibrosis was seen in 46 of 48 specimens with adherent tissue (96%). Among 61 fractured components from conical filters, 35 had extravascular penetration whereas 26 remained intravascular (11 free-floating in IVC, 15 embolized centrally), and EM revealed fracture modes of high-cycle fatigue (n = 53), overload (n = 6), and indeterminate (n = 2). Following retrieval, previously prescribed lifelong anticoagulation was discontinued in 30 of 31 patients (97%). Filter-related symptoms from IVC occlusion, component embolization, and penetration-induced abdominal pain, duodenal injury, and/or small-bowel volvulus were alleviated in all 26 cases (100%). There were no long-term complications at a mean follow-up of 371 days (range, 67-878 d). Conclusions: The risk of filter fracture increases after 408 days (ie,>1 y) of implantation and is associated with symptomatic extravascular penetration and/or intravascular embolization. Complex methods can be used to safely remove these devices, alleviate filter-related morbidity, and allow cessation of anticoagulation.
AB - Purpose: To evaluate clinical outcomes, characterize adherent tissue, and analyze inferior vena cava (IVC) filter fractures in patients undergoing complex retrieval for management of filter-related complications. To elucidate mechanisms of filter fracture by radiographic and electron microscopic (EM) evaluation. Materials and Methods: Over 2.5 years, 50 consecutive patients with fractured and/or penetrating filter components were prospectively enrolled into a single-center study. There were 19 men and 31 women (mean age, 42 y; range, 15-73 y). All patients underwent complex filter retrieval after failure of standard methods, and retrieval indications along with resultant clinical outcomes were evaluated. Specimens with adherent tissue underwent histologic analysis, and all fractured components were studied with EM. Results: Retrieval was successful in all 50 cases (mean implantation, 815 d; range, 20-2,599 d) among the following filters: G2X (n = 23),G2 (n = 9), Eclipse (n = 3), Recovery (n = 4), ALN (n = 1), Celect (n = 7), OptEase (n = 2), and Simon Nitinol (n = 1). Mean indwell time in fractured filters (n = 31) was 1,082 days, versus 408 days in nonfractured filters (n = 19; P =.00169). Neointimal hyperplasia/fibrosis was seen in 46 of 48 specimens with adherent tissue (96%). Among 61 fractured components from conical filters, 35 had extravascular penetration whereas 26 remained intravascular (11 free-floating in IVC, 15 embolized centrally), and EM revealed fracture modes of high-cycle fatigue (n = 53), overload (n = 6), and indeterminate (n = 2). Following retrieval, previously prescribed lifelong anticoagulation was discontinued in 30 of 31 patients (97%). Filter-related symptoms from IVC occlusion, component embolization, and penetration-induced abdominal pain, duodenal injury, and/or small-bowel volvulus were alleviated in all 26 cases (100%). There were no long-term complications at a mean follow-up of 371 days (range, 67-878 d). Conclusions: The risk of filter fracture increases after 408 days (ie,>1 y) of implantation and is associated with symptomatic extravascular penetration and/or intravascular embolization. Complex methods can be used to safely remove these devices, alleviate filter-related morbidity, and allow cessation of anticoagulation.
KW - EM
KW - FDA
KW - Food and Drug Administration
KW - IVC
KW - PE
KW - electron microscopy
KW - inferior vena cava
KW - pulmonary embolism
UR - http://www.scopus.com/inward/record.url?scp=84876891586&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84876891586&partnerID=8YFLogxK
U2 - 10.1016/j.jvir.2013.01.008
DO - 10.1016/j.jvir.2013.01.008
M3 - Article
C2 - 23523157
AN - SCOPUS:84876891586
SN - 1051-0443
VL - 24
SP - 622-630.e1
JO - Journal of Vascular and Interventional Radiology
JF - Journal of Vascular and Interventional Radiology
IS - 5
ER -