TY - JOUR
T1 - Biomechanics of thoracic short versus long fixation after 3-column injury
T2 - Laboratory investigation
AU - Lazaro, Bruno C.R.
AU - Deniz, Fatih Ersay
AU - Brasiliense, Leonardo B.C.
AU - Reyes, Phillip M.
AU - Sawa, Anna G.U.
AU - Theodore, Nicholas
AU - Sonntag, Volker K.H.
AU - Crawford, Neil R.
PY - 2011/2/1
Y1 - 2011/2/1
N2 - Object. Posterior screw-rod fixation for thoracic spine trauma usually involves fusion across long segments. Biomechanical data on screw-based short-segment fixation for thoracic fusion are lacking. The authors compared the effects of spanning short and long segments in the thoracic spine. Methods. Seven human spine segments (5 segments from T-2 to T-8; 2 segments from T-3 to T-9) were prepared. Pure-moment loading of 6 Nm was applied to induce flexion, extension, lateral bending, and axial rotation while 3D motion was measured optoelectronically. Normal specimens were tested, and then a wedge fracture was created on the middle vertebra after cutting the posterior ligaments. Five conditions of instrumentation were tested, as follows: Step A, 4-level fixation plus cross-link; Step B, 2-level fixation; Step C, 2-level fixation plus cross-link; Step D, 2-level fixation plus screws at fracture site (index); and Step E, 2-level fixation plus index screws plus cross-link. Results. Long-segment fixation restricted 2-level range of motion (ROM) during extension and lateral bending significantly better than the most rigid short-segment construct. Adding index screws in short-segment constructs significantly reduced ROM during flexion, lateral bending, and axial rotation (p < 0.03). A cross-link reduced axial rotation ROM (p = 0.001), not affecting other loading directions (p > 0.4). Conclusions. Thoracic short-segment fixation provides significantly less stability than long-segment fixation for the injury studied. Adding a cross-link to short fixation improved stability only during axial rotation. Adding a screw at the fracture site improved short-segment stability by an average of 25%.
AB - Object. Posterior screw-rod fixation for thoracic spine trauma usually involves fusion across long segments. Biomechanical data on screw-based short-segment fixation for thoracic fusion are lacking. The authors compared the effects of spanning short and long segments in the thoracic spine. Methods. Seven human spine segments (5 segments from T-2 to T-8; 2 segments from T-3 to T-9) were prepared. Pure-moment loading of 6 Nm was applied to induce flexion, extension, lateral bending, and axial rotation while 3D motion was measured optoelectronically. Normal specimens were tested, and then a wedge fracture was created on the middle vertebra after cutting the posterior ligaments. Five conditions of instrumentation were tested, as follows: Step A, 4-level fixation plus cross-link; Step B, 2-level fixation; Step C, 2-level fixation plus cross-link; Step D, 2-level fixation plus screws at fracture site (index); and Step E, 2-level fixation plus index screws plus cross-link. Results. Long-segment fixation restricted 2-level range of motion (ROM) during extension and lateral bending significantly better than the most rigid short-segment construct. Adding index screws in short-segment constructs significantly reduced ROM during flexion, lateral bending, and axial rotation (p < 0.03). A cross-link reduced axial rotation ROM (p = 0.001), not affecting other loading directions (p > 0.4). Conclusions. Thoracic short-segment fixation provides significantly less stability than long-segment fixation for the injury studied. Adding a cross-link to short fixation improved stability only during axial rotation. Adding a screw at the fracture site improved short-segment stability by an average of 25%.
KW - 3-column injury
KW - Pedicle screw
KW - Short-segment fixation
KW - Thoracic fixation
UR - http://www.scopus.com/inward/record.url?scp=79551707008&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=79551707008&partnerID=8YFLogxK
U2 - 10.3171/2010.10.SPINE09785
DO - 10.3171/2010.10.SPINE09785
M3 - Article
C2 - 21184641
AN - SCOPUS:79551707008
SN - 1547-5654
VL - 14
SP - 226
EP - 234
JO - Journal of Neurosurgery: Spine
JF - Journal of Neurosurgery: Spine
IS - 2
ER -