TY - GEN
T1 - Autonomous Needle Navigation in Retinal Microsurgery
T2 - 2023 IEEE International Conference on Robotics and Automation, ICRA 2023
AU - Zhang, Peiyao
AU - Kim, Ji Woong
AU - Gehlbach, Peter
AU - Iordachita, Iulian
AU - Kobilarov, Marin
N1 - Publisher Copyright:
© 2023 IEEE.
PY - 2023
Y1 - 2023
N2 - Important challenges in retinal microsurgery in-clude prolonged operating time, inadequate force feedback, and poor depth perception due to a constrained top-down view of the surgery. The introduction of robot-assisted technology could potentially deal with such challenges and improve the surgeon's performance. Motivated by such challenges, this work develops a strategy for autonomous needle navigation in retinal microsurgery aiming to achieve precise manipulation, reduced end-to-end surgery time, and enhanced safety. This is accomplished through real-time geometry estimation and chance-constrained Model Predictive Control (MPC) resulting in high positional accuracy while keeping scleral forces within a safe level. The robotic system is validated using both open-sky and intact (with lens and partial vitreous removal) ex vivo porcine eyes. The experimental results demonstrate that the generation of safe control trajectories is robust to small motions associated with head drift. The mean navigation time and scleral force for MPC navigation experiments are 7.208 s and 11.97 mN, which can be considered efficient and well within acceptable safe limits. The resulting mean errors along lateral directions of the retina are below 0.06 mm, which is below the typical hand tremor amplitude in retinal microsurgery.
AB - Important challenges in retinal microsurgery in-clude prolonged operating time, inadequate force feedback, and poor depth perception due to a constrained top-down view of the surgery. The introduction of robot-assisted technology could potentially deal with such challenges and improve the surgeon's performance. Motivated by such challenges, this work develops a strategy for autonomous needle navigation in retinal microsurgery aiming to achieve precise manipulation, reduced end-to-end surgery time, and enhanced safety. This is accomplished through real-time geometry estimation and chance-constrained Model Predictive Control (MPC) resulting in high positional accuracy while keeping scleral forces within a safe level. The robotic system is validated using both open-sky and intact (with lens and partial vitreous removal) ex vivo porcine eyes. The experimental results demonstrate that the generation of safe control trajectories is robust to small motions associated with head drift. The mean navigation time and scleral force for MPC navigation experiments are 7.208 s and 11.97 mN, which can be considered efficient and well within acceptable safe limits. The resulting mean errors along lateral directions of the retina are below 0.06 mm, which is below the typical hand tremor amplitude in retinal microsurgery.
UR - http://www.scopus.com/inward/record.url?scp=85168684184&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85168684184&partnerID=8YFLogxK
U2 - 10.1109/ICRA48891.2023.10161151
DO - 10.1109/ICRA48891.2023.10161151
M3 - Conference contribution
AN - SCOPUS:85168684184
T3 - Proceedings - IEEE International Conference on Robotics and Automation
SP - 4661
EP - 4667
BT - Proceedings - ICRA 2023
PB - Institute of Electrical and Electronics Engineers Inc.
Y2 - 29 May 2023 through 2 June 2023
ER -