Automatic vessel attenuation measurement for quality control of contrast-enhanced CT: Validation on the portal vein

Kevin McCoy, Sujay Marisetty, Dominique Tan, Corey T. Jensen, Jeffrey H. Siewerdsen, Christine B. Peterson, Moiz Ahmad

Research output: Contribution to journalArticlepeer-review

Abstract

Background: Adequate image enhancement of organs and blood vessels of interest is an important aspect of image quality in contrast-enhanced computed tomography (CT). There is a need for an objective method for evaluation of vessel contrast that can be automatically and systematically applied to large sets of CT exams. Purpose: The purpose of this work was to develop a method to automatically segment and measure attenuation Hounsfield Unit (HU) in the portal vein (PV) in contrast-enhanced abdomen CT examinations. Methods: Input CT images were processed by a vessel enhancing filter to determine candidate PV segmentations. Multiple machine learning (ML) classifiers were evaluated for classifying a segmentation as corresponding to the PV based on segmentation shape, location, and intensity features. A public data set of 82 contrast-enhanced abdomen CT examinations was used to train the method. An optimal ML classifier was selected by training and tuning on 66 out of the 82 exams (80% training split) in the public data set. The method was evaluated in terms of segmentation classification accuracy and PV attenuation measurement accuracy, compared to manually determined ground truth, on a test set of the remaining 16 exams (20% test split) held out from public data set. The method was further evaluated on a separate, independently collected test set of 21 examinations. Results: The best classifier was found to be a random forest, with a precision of 0.892 in the held-out test set to correctly identify the PV from among the input candidate segmentations. The mean absolute error of the measured PV attenuation relative to ground truth manual measurement was 13.4 HU. On the independent test set, the overall precision decreased to 0.684. However, the PV attenuation measurement remained relatively accurate with a mean absolute error of 15.2 HU. Conclusions: The method was shown to accurately measure PV attenuation over a large range of attenuation values, and was validated in an independently collected dataset. The method did not require time-consuming manual contouring to supervise training. The method may be applied to systematic quality control of contrast-enhanced CT examinations.

Original languageEnglish (US)
JournalMedical physics
DOIs
StateAccepted/In press - 2024
Externally publishedYes

Keywords

  • computed tomography
  • contrast enhancement
  • image quality
  • machine learning
  • quality control
  • random forest

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Automatic vessel attenuation measurement for quality control of contrast-enhanced CT: Validation on the portal vein'. Together they form a unique fingerprint.

Cite this