Abstract
PURPOSE: To determine what influence automatic patient-instruction (API) devices have on image quality of chest computed tomographic (CT) scans and whether the qualitative outcome justifies their routine use. MATERIALS AND METHODS: Thin-collimation CT scans of two age- and sex-matched groups of 64 patients each were evaluated prospectively for the presence of breathing artifacts and for concomitant deterioration of image quality. Breathing commands in group 1 were given with the API device and in group 2 with technologist-performed patient instruction. Cardiac motion artifacts were not evaluated. The frequency of scans repeated owing to breathing artifact was determined. RESULTS: Image quality was worse in group 1 compared with that of group 2. The percentage of scans repeated was higher with API (38%) than without API (16%). CONCLUSION: API devices cannot be recommended for thin- section CT of the thorax. The large number of scans that must be repeated leads to a considerable increase in patient irradiation, scanning time, and cost.