TY - JOUR
T1 - Augmented reality-based feedback for technician-in-the-loop C-arm repositioning
AU - Unberath, Mathias
AU - Fotouhi, Javad
AU - Hajek, Jonas
AU - Maier, Andreas
AU - Osgood, Greg
AU - Taylor, Russell
AU - Armand, Mehran
AU - Navab, Nassir
N1 - Publisher Copyright:
© 2018 Institution of Engineering and Technology.All right reserved.
PY - 2018
Y1 - 2018
N2 - Interventional C-arm imaging is crucial to percutaneous orthopedic procedures as it enables the surgeon to monitor the progress of surgery on the anatomy level. Minimally invasive interventions require repeated acquisition of X-ray images from different anatomical views to verify tool placement. Achieving and reproducing these views often comes at the cost of increased surgical time and radiation. We propose a marker-free ‘technician-in-the-loop’ Augmented Reality (AR) solution for C-arm repositioning. The X-ray technician operating the C-arm interventionally is equipped with a head-mounted display system capable of recording desired C-arm poses in 3D via an integrated infrared sensor. For C-arm repositioning to a target view, the recorded pose is restored as a virtual object and visualized in an AR environment, serving as a perceptual reference for the technician. Our proof-of-principle findings from a simulated trauma surgery indicate that the proposed system can decrease the 2.76 X-ray images required for re-aligning the scanner with an intra-operatively recorded C-arm view down to zero, suggesting substantial reductions of radiation dose. The proposed AR solution is a first step towards facilitating communication between the surgeon and the surgical staff, improving the quality of surgical image acquisition, and enabling context-aware guidance for surgery rooms of the future.
AB - Interventional C-arm imaging is crucial to percutaneous orthopedic procedures as it enables the surgeon to monitor the progress of surgery on the anatomy level. Minimally invasive interventions require repeated acquisition of X-ray images from different anatomical views to verify tool placement. Achieving and reproducing these views often comes at the cost of increased surgical time and radiation. We propose a marker-free ‘technician-in-the-loop’ Augmented Reality (AR) solution for C-arm repositioning. The X-ray technician operating the C-arm interventionally is equipped with a head-mounted display system capable of recording desired C-arm poses in 3D via an integrated infrared sensor. For C-arm repositioning to a target view, the recorded pose is restored as a virtual object and visualized in an AR environment, serving as a perceptual reference for the technician. Our proof-of-principle findings from a simulated trauma surgery indicate that the proposed system can decrease the 2.76 X-ray images required for re-aligning the scanner with an intra-operatively recorded C-arm view down to zero, suggesting substantial reductions of radiation dose. The proposed AR solution is a first step towards facilitating communication between the surgeon and the surgical staff, improving the quality of surgical image acquisition, and enabling context-aware guidance for surgery rooms of the future.
UR - http://www.scopus.com/inward/record.url?scp=85056400164&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85056400164&partnerID=8YFLogxK
U2 - 10.1049/htl.2018.5066
DO - 10.1049/htl.2018.5066
M3 - Article
C2 - 30464844
AN - SCOPUS:85056400164
SN - 2053-3713
VL - 5
SP - 143
EP - 147
JO - Healthcare Technology Letters
JF - Healthcare Technology Letters
IS - 5
ER -