N2 - Purpose: Corneal collagen cross-linking (CXL) alone cannot substantially improve refractive errors. We designed a novel corneal-shaping device, showing that a prototype applied during CXL can alter corneal curvature to a greater extent than CXL alone in a porcine eye model. Methods: The device prototype was made with flat, UV-transmissible material. Enucleated porcine eyes were deepithelialized and perfused. Preexperimental and postexperimental corneal curvature (K) measurements and Anterior Segment Optical Coherence Tomography imaging were performed. A conventional CXL protocol was followed. Six experiments of 6 treatment groups in duplicate were performed (n 12/group): (A) no CXL + no device, (B) no CXL + 30 minutes device, (C) 10 minutes, CXL + no device, (D) 10 minutes, CXL + device, (E) 30 minutes, CXL + no device, and (F) 30 minutes, CXL + device. Results: There was a significant decrease (P < 0.001) in corneal curvature after 30 minutes. CXL between group F (-1.54 ± 0.90 D) and groups E (+0.34 ± 0.53 D), B (+0.69 ± 1.02 D), or A (+0.22 ± 1.24 D). The demarcation line depth was greater in group F (34% ± 4%) than group D (30% ± 4%, P < 0.05). Groups D (-2.3% ± 21%) and F (0.24% ± 15%) had less change in corneal thickness than in control groups A (15% ± 12%, P < 0.05) and E (12% ± 11%, P < 0.05). Conclusions: Application of our corneal-shaping device during CXL significantly decreased corneal curvature compared with controls. Anterior Segment Optical Coherence Tomography imaging showed comparable extent of cross-linking in eyes treated ± device. This demonstrates that our novel device used during CXL may further promote corneal strengthening and refractive correction in patients.
AB - Purpose: Corneal collagen cross-linking (CXL) alone cannot substantially improve refractive errors. We designed a novel corneal-shaping device, showing that a prototype applied during CXL can alter corneal curvature to a greater extent than CXL alone in a porcine eye model. Methods: The device prototype was made with flat, UV-transmissible material. Enucleated porcine eyes were deepithelialized and perfused. Preexperimental and postexperimental corneal curvature (K) measurements and Anterior Segment Optical Coherence Tomography imaging were performed. A conventional CXL protocol was followed. Six experiments of 6 treatment groups in duplicate were performed (n 12/group): (A) no CXL + no device, (B) no CXL + 30 minutes device, (C) 10 minutes, CXL + no device, (D) 10 minutes, CXL + device, (E) 30 minutes, CXL + no device, and (F) 30 minutes, CXL + device. Results: There was a significant decrease (P < 0.001) in corneal curvature after 30 minutes. CXL between group F (-1.54 ± 0.90 D) and groups E (+0.34 ± 0.53 D), B (+0.69 ± 1.02 D), or A (+0.22 ± 1.24 D). The demarcation line depth was greater in group F (34% ± 4%) than group D (30% ± 4%, P < 0.05). Groups D (-2.3% ± 21%) and F (0.24% ± 15%) had less change in corneal thickness than in control groups A (15% ± 12%, P < 0.05) and E (12% ± 11%, P < 0.05). Conclusions: Application of our corneal-shaping device during CXL significantly decreased corneal curvature compared with controls. Anterior Segment Optical Coherence Tomography imaging showed comparable extent of cross-linking in eyes treated ± device. This demonstrates that our novel device used during CXL may further promote corneal strengthening and refractive correction in patients.
KW - CXL plus
KW - corneal collagen cross-linking
KW - ectasia
KW - keratoconus
UR - http://www.scopus.com/inward/record.url?scp=84951907818&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84951907818&partnerID=8YFLogxK
U2 - 10.1097/ICO.0000000000000663
DO - 10.1097/ICO.0000000000000663
M3 - Article
C2 - 26509764
AN - SCOPUS:84951907818
SN - 0277-3740
VL - 35
SP - 114
EP - 121
JO - Cornea
JF - Cornea
IS - 1
ER -