@inproceedings{7defbc4b205d4001ae6ebdf38f86fb5f,
title = "Amperometric microbiosensors based on PQQ-dependent glucose dehydrogenase towards the development of an ATP biosensor for in vitro analysis",
abstract = "An amperometric biosensor for adenosine-5′- triphosphate (ATP) was developed applying a competitive assay of two glucose converting enzymes. Hexokinase and pyrroloquinoline quinone (PQQ)-dependent glucose dehydrogenase were entrapped by pH-shift induced deposition of Resydrol at a platinum microelectrode with a diameter of 25 μm. Different mediators for the PQQ-dependent glucose deydrogenase (GDH) catalyzed reaction were used applying potentials not higher than 400 mV vs. Ag/AgCl reference electrode. The obtained biosensor is selective to ATP and sensitive to it at physiological concentrations down to the nanomolar range.",
keywords = "ATP, Hexokinase, Microbiosensor, PQQ-GDH, Resydrol",
author = "C. Weber and E. Gauda and E. Hecht and B. Mizaikoff and C. Kranz",
year = "2009",
doi = "10.1007/978-3-642-03887-7_101",
language = "English (US)",
isbn = "9783642038860",
series = "IFMBE Proceedings",
publisher = "Springer Verlag",
number = "8",
pages = "351--354",
booktitle = "World Congress on Medical Physics and Biomedical Engineering",
edition = "8",
note = "World Congress on Medical Physics and Biomedical Engineering: Micro- and Nanosystems in Medicine, Active Implants, Biosensors ; Conference date: 07-09-2009 Through 12-09-2009",
}