Additional antitachycardia pacing programming strategies further reduce unnecessary implantable cardioverter-defibrillator shocks

Boyoung Joung, Daniel R. Lexcen, Chi Keong Ching, Marc T. Silver, Jonathan P. Piccini, Laurence D. Sterns, Rafael Rabinovich, Robert A. Pickett, Shufeng Liu, Mark L. Brown, Alan Cheng

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

Background: Antitachycardia pacing (ATP) is routinely used to terminate ventricular tachyarrhythmias (VTs). However, little guidance exists on the most effective programming of ATP. Objective: This study evaluated whether additional ATP sequences are more effective in reducing implantable cardioverter-defibrillator shocks. Methods: In patients from the Shock-Less study, the number of overall shocks were compared between patients programmed to ≤3 ATP sequences (VT zone) and ≤1 ATP sequence (fast ventricular tachycardia [FVT] zone) (nominal group) and patients programmed to receive additional ATP sequences in VT (>3) or FVT (>1) zones. Results: Of the 4112 patients (15% receiving secondary prevention; 77% men; mean age 65.9 ± 12.6 years), 1532 patients (37%) were programmed with additional ATP sequences (1025 with >3 ATP sequences in the VT zone; 699 patients with >1 ATP sequence in the FVT zone). Over a mean follow-up period of 19.6 ± 10.7 months, 4359 VT/FVT episodes occurred in 591 patients. Compared with the nominal group, in patients with additional ATP programming, there was a 39% reduction in the number of shocked VT episodes (0.46 episodes per patient-year vs 0.28 episodes per patient-year; incidence rate ratio [IRR] 0.61; P < .001) and a 44% reduction in the number of shocked FVT episodes (0.83 episodes per patient-year vs 0.47 episodes per patient-year; IRR 0.56; P < .001). The reduction in shocked VT episodes was observed in both primary (IRR 0.68; 95% confidence interval 0.51–0.90; P = .007) and secondary (IRR 0.51; 95% confidence interval 0.35–0.72; P < .001) prevention patients. Conclusion: Programming more than the nominal number of ATP sequences in both the VT and FVT zones is associated with a lower occurrence of implantable cardioverter-defibrillator shocks in clinical practice.

Original languageEnglish (US)
Pages (from-to)98-105
Number of pages8
JournalHeart Rhythm
Volume17
Issue number1
DOIs
StatePublished - Jan 2020
Externally publishedYes

Keywords

  • Antitachycardia pacing
  • Implantable cardioverter-defibrillator
  • Primary prevention
  • Secondary prevention
  • Shock

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Physiology (medical)

Fingerprint

Dive into the research topics of 'Additional antitachycardia pacing programming strategies further reduce unnecessary implantable cardioverter-defibrillator shocks'. Together they form a unique fingerprint.

Cite this