Abstract
A research program for investigating the responsiveness of infants at high risk for developing schizophrenia is described briefly. Two measures of reactivity to brief auditory stimuli, the evoked heart-rate response and the evoked potential of the brain, are recorded for analysis by the PDP-12 computer. An interface to the PDP-12, called the R peak detector and R-R interval counter, is used for reducing the EKG to R-R intervals, with an accuracy of ± 1 msec. The current usage of the PDP-12 and its IBM-compatible tape transports is discussed, and proposed developments for on-line statistical analyses of the physiological data and real-time control of the experiment are presented.
Original language | English (US) |
---|---|
Pages (from-to) | 87-88 |
Number of pages | 2 |
Journal | Behavior Research Methods & Instrumentation |
Volume | 3 |
Issue number | 2 |
DOIs | |
State | Published - Mar 1971 |
Externally published | Yes |
ASJC Scopus subject areas
- Experimental and Cognitive Psychology
- Psychology (miscellaneous)
- Psychology(all)