N2 - Material decomposition in CT has the potential to reduce artifacts and improve quantitative accuracy by utilizing spectral models and multi-energy scans. In this work we present a novel Model-Based Material Decomposition (MBMD) method based on an existing iterative reconstruction algorithm derived from a general non-linear forward model. A digital water phantom with inserts containing different concentrations of calcium was scanned on a kV switching system. We used the presented method to simultaneously reconstruct water and calcium material density images, and compared the results to an image domain and a projection domain decomposition method. When switching voltage every other frame, MBMD resulted in more accurate water and calcium concentration values than the image domain decomposition method, and was just as accurate as the projection domain decomposition method. In a second, slower, kV switching scheme (changing voltage every ten frames) which precluded the use of traditional projection domain based methods, MBMD continued to produce quantitatively accurate reconstructions. Finally, we present a preliminary study applying MBMD to a water phantom containing vials of different concentrations of K2HPO4 which was scanned on a cone-beam CT test bench. Both the fast and slow (emulated) kV switching schemes resulted in similar reconstructions, indicating MBMD's robustness to challenging acquisition schemes. Additionally, the K2HPO4 concentration ratios between the vials were accurately represented in the reconstructed K2HPO4 density image.
AB - Material decomposition in CT has the potential to reduce artifacts and improve quantitative accuracy by utilizing spectral models and multi-energy scans. In this work we present a novel Model-Based Material Decomposition (MBMD) method based on an existing iterative reconstruction algorithm derived from a general non-linear forward model. A digital water phantom with inserts containing different concentrations of calcium was scanned on a kV switching system. We used the presented method to simultaneously reconstruct water and calcium material density images, and compared the results to an image domain and a projection domain decomposition method. When switching voltage every other frame, MBMD resulted in more accurate water and calcium concentration values than the image domain decomposition method, and was just as accurate as the projection domain decomposition method. In a second, slower, kV switching scheme (changing voltage every ten frames) which precluded the use of traditional projection domain based methods, MBMD continued to produce quantitatively accurate reconstructions. Finally, we present a preliminary study applying MBMD to a water phantom containing vials of different concentrations of K2HPO4 which was scanned on a cone-beam CT test bench. Both the fast and slow (emulated) kV switching schemes resulted in similar reconstructions, indicating MBMD's robustness to challenging acquisition schemes. Additionally, the K2HPO4 concentration ratios between the vials were accurately represented in the reconstructed K2HPO4 density image.
UR - http://www.scopus.com/inward/record.url?scp=85049235093&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85049235093&partnerID=8YFLogxK
U2 - 10.1117/12.2293776
DO - 10.1117/12.2293776
M3 - Conference contribution
C2 - 29643571
AN - SCOPUS:85049235093
T3 - Progress in Biomedical Optics and Imaging - Proceedings of SPIE
BT - Medical Imaging 2018
A2 - Schmidt, Taly Gilat
A2 - Chen, Guang-Hong
A2 - Lo, Joseph Y.
PB - SPIE
T2 - Medical Imaging 2018: Physics of Medical Imaging
Y2 - 12 February 2018 through 15 February 2018
ER -