TY - GEN
T1 - A computer assisted method for guide-wire and catheter evaluation
AU - Patriciu, A.
AU - Mazilu, D.
AU - Petrisor, D.
AU - Bagga, H. S.
AU - Kavoussi, L.
AU - Stoianovici, D.
PY - 2005
Y1 - 2005
N2 - We developed a method for testing guide wires and catheters that realistically evaluates the forces applied to anatomical structures by these instruments during urological procedures. The placement of guide wires and catheters to gain access to the upper urinary tract can induce undesirable stresses on the tissues. Previous studies have characterized wire/catheter performances base on their physical properties, such as stiffness and friction coefficient. However, the results of these studies do not directly quantify their effect on the tissues. Additionally, individual physical properties do not entirely characterize the behavior of the wire/catheter ensemble. Our model utilizes a Computer-controlled test stand that simulates the urological environment by including a tortuous path and a stone obstruction. Experimental results indicate that the method shows significant promise in reflecting wire/catheter performance data in congruence with reliable real-life measures of stress upon relevant anatomical structures. Furthermore, due to the modularity of the approach, the model can bei easily reconfigured to simulate Environments from other medical fields.
AB - We developed a method for testing guide wires and catheters that realistically evaluates the forces applied to anatomical structures by these instruments during urological procedures. The placement of guide wires and catheters to gain access to the upper urinary tract can induce undesirable stresses on the tissues. Previous studies have characterized wire/catheter performances base on their physical properties, such as stiffness and friction coefficient. However, the results of these studies do not directly quantify their effect on the tissues. Additionally, individual physical properties do not entirely characterize the behavior of the wire/catheter ensemble. Our model utilizes a Computer-controlled test stand that simulates the urological environment by including a tortuous path and a stone obstruction. Experimental results indicate that the method shows significant promise in reflecting wire/catheter performance data in congruence with reliable real-life measures of stress upon relevant anatomical structures. Furthermore, due to the modularity of the approach, the model can bei easily reconfigured to simulate Environments from other medical fields.
UR - http://www.scopus.com/inward/record.url?scp=33846903343&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33846903343&partnerID=8YFLogxK
M3 - Conference contribution
AN - SCOPUS:33846903343
SN - 0780387406
SN - 9780780387409
T3 - Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings
SP - 436
EP - 439
BT - Proceedings of the 2005 27th Annual International Conference of the Engineering in Medicine and Biology Society, IEEE-EMBS 2005
T2 - 2005 27th Annual International Conference of the Engineering in Medicine and Biology Society, IEEE-EMBS 2005
Y2 - 1 September 2005 through 4 September 2005
ER -