TY - JOUR
T1 - A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform
AU - Tryggestad, E.
AU - Armour, M.
AU - Iordachita, I.
AU - Verhaegen, F.
AU - Wong, J. W.
PY - 2009
Y1 - 2009
N2 - Our group has constructed the small animal radiation research platform (SARRP) for delivering focal, kilo-voltage radiation to targets in small animals under robotic control using cone-beam CT guidance. The present work was undertaken to support the SARRP's treatment planning capabilities. We have devised a comprehensive system for characterizing the radiation dosimetry in water for the SARRP and have developed a Monte Carlo dose engine with the intent of reproducing these measured results. We find that the SARRP provides sufficient therapeutic dose rates ranging from 102 to 228 cGy min-1 at 1 cm depth for the available set of high-precision beams ranging from 0.5 to 5 mm in size. In terms of depth-dose, the mean of the absolute percentage differences between the Monte Carlo calculations and measurement is 3.4% over the full range of sampled depths spanning 0.5-7.2 cm for the 3 and 5 mm beams. The measured and computed profiles for these beams agree well overall; of note, good agreement is observed in the profile tails. Especially for the smallest 0.5 and 1 mm beams, including a more realistic description of the effective x-ray source into the Monte Carlo model may be important.
AB - Our group has constructed the small animal radiation research platform (SARRP) for delivering focal, kilo-voltage radiation to targets in small animals under robotic control using cone-beam CT guidance. The present work was undertaken to support the SARRP's treatment planning capabilities. We have devised a comprehensive system for characterizing the radiation dosimetry in water for the SARRP and have developed a Monte Carlo dose engine with the intent of reproducing these measured results. We find that the SARRP provides sufficient therapeutic dose rates ranging from 102 to 228 cGy min-1 at 1 cm depth for the available set of high-precision beams ranging from 0.5 to 5 mm in size. In terms of depth-dose, the mean of the absolute percentage differences between the Monte Carlo calculations and measurement is 3.4% over the full range of sampled depths spanning 0.5-7.2 cm for the 3 and 5 mm beams. The measured and computed profiles for these beams agree well overall; of note, good agreement is observed in the profile tails. Especially for the smallest 0.5 and 1 mm beams, including a more realistic description of the effective x-ray source into the Monte Carlo model may be important.
UR - http://www.scopus.com/inward/record.url?scp=70350496336&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=70350496336&partnerID=8YFLogxK
U2 - 10.1088/0031-9155/54/17/017
DO - 10.1088/0031-9155/54/17/017
M3 - Article
C2 - 19687532
AN - SCOPUS:70350496336
SN - 0031-9155
VL - 54
SP - 5341
EP - 5357
JO - Physics in medicine and biology
JF - Physics in medicine and biology
IS - 17
ER -