A cadaveric precision and accuracy analysis of augmented reality–mediated percutaneous pedicle implant insertion

Camilo A. Molina, Frank M. Phillips, Matthew W. Colman, Wilson Z. Ray, Majid Khan, Emanuele Orru', Kornelis Poelstra, Larry Khoo

Research output: Contribution to journalArticlepeer-review

Abstract

OBJECTIVE Augmented reality–mediated spine surgery (ARMSS) is a minimally invasive novel technology that has the potential to increase the efficiency, accuracy, and safety of conventional percutaneous pedicle screw insertion methods. Visual 3D spinal anatomical and 2D navigation images are directly projected onto the operator’s retina and superimposed over the surgical field, eliminating field of vision and attention shift to a remote display. The objective of this cadaveric study was to assess the accuracy and precision of percutaneous ARMSS pedicle implant insertion. METHODS Instrumentation was placed in 5 cadaveric torsos via ARMSS with the xvision augmented reality head-mounted display (AR-HMD) platform at levels ranging from T5 to S1 for a total of 113 total implants (93 pedicle screws and 20 Jamshidi needles). Postprocedural CT scans were graded by two independent neuroradiologists using the Gertzbein-Robbins scale (grades A–E) for clinical accuracy. Technical precision was calculated using superimposition analysis employing the Medical Image Interaction Toolkit to yield angular trajectory (°) and linear screw tip (mm) deviation from the virtual pedicle screw position compared with the actual pedicle screw position on postprocedural CT imaging. RESULTS The overall implant insertion clinical accuracy achieved was 99.1%. Lumbosacral and thoracic clinical accuracies were 100% and 98.2%, respectively. Specifically, among all implants inserted, 112 were noted to be Gertzbein-Robbins grade A or B (99.12%), with only 1 medial Gertzbein-Robbins grade C breach (> 2-mm pedicle breach) in a thoracic pedicle at T9. Precision analysis of the inserted pedicle screws yielded a mean screw tip linear deviation of 1.98 mm (99% CI 1.74–2.22 mm) and a mean angular error of 1.29° (99% CI 1.11°–1.46°) from the projected trajectory. These data compare favorably with data from existing navigation platforms and regulatory precision requirements mandating that linear and angular deviation be less than 3 mm (p < 0.01) and 3° (p < 0.01), respectively. CONCLUSIONS Percutaneous ARMSS pedicle implant insertion is a technically feasible, accurate, and highly precise method.

Original languageEnglish (US)
Pages (from-to)316-324
Number of pages9
JournalJournal of Neurosurgery: Spine
Volume34
Issue number2
DOIs
StatePublished - Feb 2021

Keywords

  • Augmented reality–assisted surgery
  • Cadaveric trial
  • Computer-navigated surgery
  • Minimally invasive spine surgery
  • Mixed reality–assisted surgery
  • Neuronavigation
  • Percutaneous pedicle screw insertion
  • Robotic spine surgery
  • Surgical technique

ASJC Scopus subject areas

  • Surgery
  • Neurology
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'A cadaveric precision and accuracy analysis of augmented reality–mediated percutaneous pedicle implant insertion'. Together they form a unique fingerprint.

Cite this